Calculate concentration, absorbance, or molar absorptivity using Beer-Lambert law (A = εlc). Includes nucleic acid quantification and common chromophore database.
Load example Beer-Lambert data to see the full workflow
When to use
Determine analyte concentration from a known absorbance reading and extinction coefficient
Predict expected absorbance before running a spectrophotometry experiment
Look up molar absorptivity values for common chromophores, dyes, and proteins
Quantify DNA or RNA concentration from A260/A280 readings
Batch-convert absorbance readings at multiple wavelengths to concentrations
Do not use for
For turbid or scattering samples without prior clarification — absorbance includes apparent scatter
When absorbance exceeds 2.0 — dilute first and re-measure
As a substitute for a validated standard curve in regulated assays
Always blank your spectrophotometer with the correct solvent
The blank must match the sample matrix (solvent, buffer, cuvette type). A mismatched blank introduces systematic offset in every reading. For nucleic acids, blank with the elution buffer, not water.
Extinction coefficients are wavelength- and solvent-specific
Literature ε values are measured at a specific wavelength in a specific solvent. Using an ε measured in water for a sample in DMSO, or at 280 nm instead of 260 nm, will give incorrect concentrations.
Path length is NOT always 1 cm
Microvolume instruments like NanoDrop use 0.05–1 mm path lengths. Microplate readers have variable path lengths depending on well volume. Always verify and enter the actual path length.
The A260/A280 ratio is not a purity assay
While A260/A280 ≈ 1.8 suggests pure DNA, contamination with co-absorbing substances at both wavelengths (e.g., some phenolic compounds) can yield a “normal” ratio. Gel electrophoresis or fluorometric quantification provides better purity assessment.
1
Method
Direct algebraic solution of the Beer-Lambert equation (A=εlc). Nucleic acid quantification uses consensus conversion factors (dsDNA: 50 µg/mL per A260, ssDNA: 33, RNA: 40). Extinction coefficients sourced from primary literature and standard biochemistry references.
2
Published
Published 2026-04-05. Calculations support planning and documentation; verify procurement decisions against manufacturer specifications or institutional SOPs.
3
How to cite
How to Cite
ConductScience Beer-Lambert Calculator. ConductScience, Inc. 2026. Available at: https://conductscience.com/tools/beer-lambert-calculator
Beer A. Bestimmung der Absorption des rothen Lichts in farbigen Flüssigkeiten. Ann Phys. 1852;162(5):78–88.
The Beer-Lambert law relates the attenuation of light to the properties of the material through which it travels:
A=ε×l×c
Where:
• A = Absorbance (unitless, also called optical density)
• ε = Molar absorptivity (L⋅mol−1⋅cm−1)
• l = Path length through the sample (cm)
• c = Molar concentration (mol/L)
Absorbance is defined as A = –log₁₀(T), where T is transmittance (I/I₀). This means A = 1 corresponds to 10% transmittance, and A = 2 corresponds to 1% transmittance. The linear relationship holds for dilute, homogeneous solutions of non-interacting chromophores.
Common Pitfalls in Spectrophotometry
Several factors can cause deviations from Beer-Lambert linearity:
• High concentration: Molecular interactions and aggregation change the effective ε
• Stray light: Instrument imperfections add a constant signal, compressing high-absorbance readings
• Scattering: Turbid samples scatter light, mimicking absorbance
• Chemical equilibria: pH-dependent speciation can shift with concentration (e.g., indicator dyes)
• Fluorescence: Re-emitted photons reach the detector, reducing apparent absorbance
• Wrong wavelength: Measuring off-peak reduces sensitivity and may introduce nonlinearity
Best practice: run a standard curve at your exact conditions and verify linearity before quantifying unknowns.
Frequently asked
What is the Beer-Lambert law?
The Beer-Lambert law (also called Beer’s law) describes the linear relationship between absorbance and concentration of an absorbing species in solution: A=εlc, where A is absorbance (unitless), ε is the molar extinction coefficient (L⋅mol−1⋅cm−1), l is the optical path length (cm), and c is concentration (mol/L). It is the foundation of quantitative UV-Vis spectrophotometry.
When does the Beer-Lambert law fail?
The law assumes dilute solutions with non-interacting chromophores. It breaks down at high concentrations (typically A > 2) due to molecular interactions, aggregation, and changes in refractive index. Scattering samples, fluorescent compounds, and chemical equilibria that shift with concentration also cause deviations. Always verify linearity with a standard curve.
What is molar absorptivity (extinction coefficient)?
Molar absorptivity (ε), also called the molar extinction coefficient, is a measure of how strongly a chemical species absorbs light at a given wavelength. It is an intrinsic property of the molecule in a specific solvent, expressed in L⋅mol−1⋅cm−1. Higher ε values mean stronger absorption and thus greater analytical sensitivity.
Why is absorbance above 2 unreliable?
An absorbance of 2 means only 1% of light reaches the detector (10^-2 = 0.01). At A > 2, the detected signal is so small that electronic noise dominates, and stray light in the instrument causes systematic underestimation of the true absorbance. Most spectrophotometers are accurate only up to A ≈ 1.5–2.0. Dilute the sample to bring readings into the linear range (A = 0.1–1.0).
How does the A260 method work for DNA/RNA quantification?
Nucleic acids absorb UV light strongly at 260 nm due to their purine and pyrimidine bases. By convention, an A260 of 1.0 in a 1 cm cuvette corresponds to approximately 50 µg/mL of dsDNA, 33 µg/mL of ssDNA, or 40 µg/mL of RNA. The A260/A280 ratio assesses purity: ~1.8 for pure DNA, ~2.0 for pure RNA. Ratios below 1.7 suggest protein or phenol contamination.
How do I correct for path length differences?
Standard cuvettes have a 1 cm path length, but microvolume instruments (e.g., NanoDrop) use much shorter paths (0.05–1 mm). The Beer-Lambert law accounts for this: simply enter the correct path length value. For microplate readers, path length varies with well volume and must be corrected — many instruments offer an automatic path length correction feature using the A977/A900 water absorbance ratio.