AZtec projects reach the data size limit
<p>Ten Ti-6Al-4V samples were mounted on a multi-sample stage for EBSD on a Thermo Fisher Apreo SEM equipped with an Oxford Instruments' Symmetry 2 detector at the University of Manchester.</p> <p>In project <em>multi-sample_1</em>, AZtec reported a saving error when scanning the fifth sample and stopped with 5646 frames saved (.oip~4GB). It is able to montage and export the maps, but any edit on the .oip file cannot be saved.</p> <p>In project <em>multi-sample_2,</em> we restarted the scan on the rest of the samples and completed with 5601 frames. The .oip is 3.97GB, which almost reaches the size limit. No error was reported during the scanning, and the .oip file is still editable. </p>