3D ED / MicroED data of BI-3812
<p>The sample was obtained from Boehringer Ingelheim within a frame of opnme project <a href="https://www.opnme.com">https://www.opnme.com</a></p> <p>Electron diffraction tilt series were recorded with GLACIOS TEM (Thermo Fisher) at liquid nitrogen temperature. Measurements were performed using the EPU-D (Thermo Fisher) module in continuous rotation mode with 200 kV electrons (wavelength of 0.0251 Å). Tilt series were collected within ±60° goniometer tilt range with 1° frame interval and rotation speed of 1° per second. The data were collected using an electron dose rate of 6.016 e/nm<sup>2</sup>s. The electron beam size used for diffraction data collection was 1.2 µm. The pixel size within diffraction patterns is 0.001075 Å<sup>-1</sup>.</p>
ShareScore
32/100
Overall dataset sharing score
Score breakdown
These five areas show where the dataset supports — or may limit — practical reuse.
- Stewardship
- 8
- Harmonization
- 4
- Access
- 16
- Reuse readiness
- 4
- Engagement
- 0
Topics
Dataset details
- Repository
- zenodo
- Published
- April 2, 2023
- License
- cc-by-4.0
- DOI
- 10.5281/zenodo.7794778