Dual-View ICP-OES Spectrometer with Vertical Torch
Vertical-torch ICP-OES with full-spectrum real-time correction and <1% RSD over 8 hours.

Louise Corscadden, PhD
Director of Science · ConductScience
Ask Louise about Dual-View ICP-OES Spectrometer with Vertical Torch fit, setup, configuration, or quote prep.
Already working with us? Sign in to connect this with My Scientist.
Key Specifications
Full details →- Supply voltage
- 220V AC ± 10%
- Ambient temperature
- 10-30℃
- Relative humidity
- 20-80%RH
- RF power range
- 500 W - 1,600 W
- Standby RF power
- 500 W
- Standby argon consumption
- ≤5 L/min
An optical emission spectrometer whose vertical torch and dual-view optics let a single instrument cover both trace and major-element concentrations. Full-spectrum real-time correction holds calibration across a working day, and the solid-state RF generator tolerates high dissolved-solids loading.
Key features
- A new generation of vertical torch dual observation technology
- Patented self-excited all-solid-state RF power supply
- The large area array ECCD sensor improves the sensitivity and spectral range of the instrument.
- 8-hour stability RSD < 1%
- All-solid-state constant temperature two-dimensional optical system.
- Full-spectrum real-time correction system, ready to be used after it is booted.
- Less refraction and reflection to ensure less loss of UV light intensity.
- Atomizing gas, cooling gas, and auxiliary gas, under precise control.
Applications
Built for alloy, geochemical and high-salinity testing laboratories.
- The ICP-OES supports the analysis of rare earths with requirements for full-spectrum high-resolution, and the analysis of Cl/Br elements over the deep ultraviolet waveband.
- Applications on vehicles are supported in a stable and reliable way.
Choosing the right configuration
Match the sample matrix and target elements to the required detection limits, sample-introduction hardware, gas supply, exhaust, cooling, and data-system configuration. ConductScience confirms these items against the proposed method before the system is quoted.
Configurations
- Used for the dual view with the vertical torch, and applicable to various special analysis fields with high complexity and high sensitivity requirements, such as the various high-end analysis reuirements of hich-purity materials
- Used for the radial view with the vertical torch, and applicable to the general analysis demand of the complex matrix, such as the various analysis fields of alloy, high salinity, etc.
- With full spectrum and high resolution; applicable to the rare earth industry
- Features: Bidirectional observation with vertical torch. Applicable field: Applicable to areas with high sensitivity requirements, and providing a relatively wide range of versatility.
- Features: Bidirectional observation with vertical torch. Applicable field: Suitable for applications with complex matrices, such as metals, oil products, geological and mineral samples, and etc.
Features & Benefits
Supply voltage
- 220V AC ± 10%
Ambient temperature
- 10-30℃
Relative humidity
- 20-80%RH
RF power range
- 500 W - 1,600 W
Standby RF power
- 500 W
Standby argon consumption
- ≤5 L/min
Plasma observation modes
- axial, radial, bi-directional and simultaneous bi-directional
RF frequency
- 27.12 MHz
RF coupling efficiency
- >80%
RF power stability
- ≤0.1%
RF frequency stability
- ≤0.01%
Optical-chamber temperature
- 36℃±0.1℃
Wavelength range
- 160 nm - 900 nm
Monochromator focal length
- 380 mm
Echelle grating
- 87 L/mm, 63.5° blazing
Stray light
- ≤2.0 mg/L (10,000 mg/L Ca solution at As 188.980 nm)
Optical resolution (FWHM)
- ≤7 pm @ 200 nm
Detector array
- 1,024*1,024 pixels
Detector imaging area
- 25.4 mm×25.4 mm
Detector pixel size
- 24 um × 24 um
Peristaltic pump
- 12-roller 4-channel; 0 rpm to 125 rpm
Gas-control capacity
- up to 5 channels
Gas-flow control accuracy
- 0.01 L/min
Analysis speed
- about 200 spectral lines per minute
Sample consumption
- 2 ml
Spectral linear dynamic range
- ≥10⁵
Precision
- RSD ≤0.5% after ten measurements of 1 ppm or 10 ppm multi-element standard
Eight-hour stability
- RSD ≤1%
Detection limits
- Zn 0.5; Ni 1; Mn 0.5; Cr 1; Cu 1; Ba 0.1 ug/L
Preheating time
- shorter than 5 minutes
What does this instrument measure, and who is it for?
It measures major and trace elements across the same sample without changing the plasma view. It is built for alloy, geochemical and high-salinity testing laboratories. It is a benchtop instrument intended for continuous laboratory duty, with a plumbed gas supply and extraction over the source.
What measurement principle does it use?
Atoms excited in an argon plasma emit light at element-characteristic wavelengths. A vertical torch read from two viewing geometries covers trace and major concentrations on one instrument.
Which specifications are confirmed?
The specification table lists thirty confirmed ordering parameters for this instrument. Use them to compare the operating range, utilities, and analytical performance required by the method. They include Supply voltage at 220V AC ± 10%, Ambient temperature at 10-30℃ and Relative humidity at 20-80%RH.
What should be confirmed before ordering?
Confirm argon supply and purity, exhaust extraction over the torch or spark stand, single-phase power and breaker rating, bench load rating, and cooling water where the design requires it. These are configured and quoted per installation, so raise anything unusual about the site during quote review rather than after delivery.
What else is needed to run it?
A microwave digestion system, unless your samples already arrive in solution. Elemental analysis is only as good as the dissolution that preceded it, and an incomplete digest reports a low result with no indication that anything went wrong.
How is it supplied?
This instrument is configured and quoted per installation rather than sold at a list price. Request a quote and we will confirm the configuration, the site requirements and the lead time before anything is ordered.
Have a question about this product?
Have a question? Just ask.
Send it over and we'll email you a personalized answer, with no call or scheduling.
Prefer to talk it through?