Key Specifications
Full details →- Optical System
- Optical chamber with argon self-purification system
- Optical System - Chamber
- Optical chamber with argon self-purification system
- Optical System - Wavelength Range
- 174~520 nm
- Source
- Programmable full digital pulse source
- Source - Ignition
- High energy pre-ignition technology, ignition pulse: 1~14Kv
- Source - Discharge
- Maximum discharge frequency: 1000Hz, Maximum discharge current: 400A
A spark optical emission spectrometer for solid metal analysis. A programmable full-digital pulse source and a Paschen–Runge optic with CMOS detectors return a certified-grade result in seconds, and the argon self-purification system keeps the optical chamber stable between campaigns.
Key features
- Stability upgrade
- Ease of use upgrade
- Adopt scientific research level CMOS detector with better performance
- Adopt stable and reliable programmable full digital pulse optic source
- Lower Noise: Scientific research grade photosensitive element, lower noise, strong anti-interference, stability upgrade of anti-halo technology
- Anti-Interference: detector has high integration, which can avoid introducing noise into external circuits
- Fast Reading Speed: Fast reading speed: OEO (Optimal Element Oriented) technology is adopted, and pixel signals are read separately to achieve parameter optimization design.
- High UV Response: Higher UV response sensitivity, no need for coating, achieving non-metallic element (N, C, S, P) analysis, with better results
Applications
Built for foundries, metal QC and incoming-inspection laboratories.
- Casting
- Aerospace
- Precision Manufacturing
- Automobile Manufacturing
- Rail Transit
Choosing the right configuration
Match the sample matrix and target elements to the required detection limits, sample-introduction hardware, gas supply, exhaust, cooling, and data-system configuration. ConductScience confirms these items against the proposed method before the system is quoted.
Features & Benefits
Optical System
- Optical chamber with argon self-purification system
Optical System - Chamber
- Optical chamber with argon self-purification system
Optical System - Wavelength Range
- 174~520 nm
Source
- Programmable full digital pulse source
Source - Ignition
- High energy pre-ignition technology, ignition pulse: 1~14Kv
Source - Discharge
- Maximum discharge frequency: 1000Hz, Maximum discharge current: 400A
Source - Excitation pulse
- Spark excitation pulse: 20~230V, Arc excitation pulse: 20~60 V
Power supply voltage
- (220 ± 20) V AC, (50 ± 1) Hz, single power supply with protective grounding
Power
- Maximum excitation frequency: 400W, Average standby power: 50W
Environmental requirements
- Working temperature: 10~30 ℃, Storage temperature: 0~45 ℃, Working humidity: 20~80% RH
Spark Stand
- Opened sample operation platform, which can be equipped with various sample adapters for special applications to meet different small sample analysis needs
Spark Stand - Argon flushing
- Optimized argon flushing design, better argon saving and spark stability
Spark Stand - Maintenance
- Easy to clean and maintain, intelligent reminder of maintenance cycle
Aperture of spark stand
- 13mm (optional: 6mm)
Argon
- Burning flow is about 3.5L/min, maintaining and standby flow is about 0.1L/min
Argon flow
- Burning flow is about 3.5L/min, maintaining and standby flow is about 0.1L/min
Dimensions and weight
- Length: 818mm, width: 590, height: 396mm weight: about 70Kg
EMC
- IEC 610004-2, 61000-4-4, 61000-4-5
Model F
- None
Model N
- Suitable for all matrix analysis and can meet the matrix analysis requirements such as Fe, Al, Cu, Ni, and Zn
Model S
- Suitable for nonferrous metal analysis and meet the matrix analysis requirements such as Al, Cu, Zn, and Mg
Optical System (Wavelength Range)
- Wavelength Range: 174~520 nm
Source (shared)
- Programmable full digital pulse source; High energy pre-ignition technology, ignition pulse: 1~14Kv; Maximum discharge frequency: 1000Hz, Maximum discharge current: 400A; Spark excitation pulse: 20~230V, Arc excitation pulse: 20~60 V
Power supply and environmental requirements (shared)
- Power supply voltage: (220 ± 20) V AC, (50 ± 1) Hz, single power supply with protective grounding; Maximum excitation frequency: 400W, Average standby power: 50W; Working temperature: 10~30 ℃, Storage temperature: 0~45 ℃, Working humidity: 20~80% RH
Spark Stand (shared)
- Opened sample operation platform, which can be equipped with various sample adapters for special applications to meet different small sample analysis needs; Optimized argon flushing design, better argon saving and spark stability; Easy to clean and maintain, intelligent reminder of maintenance cycle; Aperture of spark stand: 13mm (optional: 6mm)
Argon (shared purity/pressure)
- Purity: 99.999%, Pressure: 0.5MPa
EMC (shared)
- IEC 610004-2, 61000-4-4, 61000-4-5
What does this instrument measure, and who is it for?
It measures the elemental composition of solid metal samples, directly and without dissolution. It is built for foundries, metal QC and incoming-inspection laboratories. It is a benchtop instrument intended for continuous laboratory duty, with a plumbed gas supply and extraction over the source.
What measurement principle does it use?
A programmable spark excites the solid surface directly; a Paschen-Runge optic with CMOS detectors reads the emission. No dissolution, no reagents, result in seconds.
Which specifications are confirmed?
The specification table lists twenty-seven confirmed ordering parameters for this instrument. Use them to compare the operating range, utilities, and analytical performance required by the method. They include Optical System - Wavelength Range at 174~520 nm, Aperture of spark stand at 13mm (optional: 6mm) and EMC at IEC 610004-2, 61000-4-4, 61000-4-5.
What should be confirmed before ordering?
Confirm argon supply and purity, exhaust extraction over the torch or spark stand, single-phase power and breaker rating, bench load rating, and cooling water where the design requires it. These are configured and quoted per installation, so raise anything unusual about the site during quote review rather than after delivery.
What else is needed to run it?
A microwave digestion system, unless your samples already arrive in solution. Elemental analysis is only as good as the dissolution that preceded it, and an incomplete digest reports a low result with no indication that anything went wrong.
How is it supplied?
This instrument is configured and quoted per installation rather than sold at a list price. Request a quote and we will confirm the configuration, the site requirements and the lead time before anything is ordered.
Have a question about this product?
Have a question? Just ask.
Send it over and we'll email you a personalized answer, with no call or scheduling.
Prefer to talk it through?