Research data · Cold storage

Energy use of lab freezers and refrigerators

On EPA's ENERGY STAR list, the median ultra-low temperature freezer uses 8.72 kWh a day at -80 °C and 5.98 kWh at -70 °C.

Model by model, -70 °C in place of -80 °C saves a median of 30% of a freezer's daily use. The saving runs from 9.5% to 40.1%.

Daily energy use of each ultra-low temperature freezer on EPA's list, at both test temperatures

Each pair of dots is one model, and the line between them is its saving at -70 °C (median 30%).

  • At -80 °C
  • At -70 °C
  • Median at -80 °C, 8.72 kWh
  • Median at -70 °C, 5.98 kWh
Every freezer's daily use
Daily energy use of each freezer at each setting, in order of volume. Some models report the same results.
Volume (liters)Volume (cubic feet)kWh a day at -80 °CkWh a day at -70 °CSaving
42114.878.015.8626.8%
42114.878.015.8626.8%
52418.55.434.3320.3%
52418.55.434.3320.3%
52718.66.845.4420.5%
53018.77.796.2719.5%
53018.78.445.9429.6%
53018.79.586.2834.4%
53118.757.145.5622.1%
53218.787.566.2018%
54719.318.715.7833.6%
54719.318.715.7833.6%
54919.398.445.6732.8%
54919.398.445.6732.8%
54919.398.445.6732.8%
56419.938.375.4934.4%
56419.938.375.4934.4%
58120.524.933.4330.4%
58120.524.933.4330.4%
58120.526.634.7228.8%
58120.526.634.7228.8%
58120.528.275.4234.5%
58120.528.275.4234.5%
58120.528.275.4234.5%
68224.098.725.9032.3%
68224.098.725.9032.3%
68224.098.725.9032.3%
68224.099.065.9834%
68224.099.065.9834%
70224.798.325.2437%
70224.798.325.2437%
71925.44.713.6622.3%
71925.44.713.6622.3%
72025.429.666.2835%
72025.429.666.2835%
72025.429.666.2835%
72025.4210.297.2829.3%
72025.4210.297.2829.3%
72025.4210.297.2829.3%
72925.749.006.7225.3%
73325.96.014.8918.6%
73325.96.014.8918.6%
73325.98.577.0018.3%
73425.938.826.3428.1%
736268.356.5122%
736268.356.5122%
75126.5310.167.8822.4%
77927.59.397.4021.2%
77927.59.957.2726.9%
79127.927.366.0717.5%
79428.046.705.5816.7%
81528.799.436.1634.7%
81528.799.436.1634.7%
81528.799.436.1634.7%
81528.7910.448.0223.2%
81628.828.785.9632.1%
81628.828.785.9632.1%
81628.828.785.9632.1%
82529.145.984.4325.9%
82529.145.984.4325.9%
82529.1411.827.7434.5%
82529.1411.827.7434.5%
82929.279.576.8628.3%
82929.279.576.8628.3%
82929.279.576.8628.3%
83929.6210.546.3140.1%
83929.6210.546.3140.1%
85230.110.137.5825.2%
86130.410.628.9615.6%
93733.0910.859.829.5%
95033.569.676.6231.5%
95033.569.676.6231.5%
95933.859.306.1733.7%
95933.859.306.1733.7%
96033.95.604.3023.2%
96033.95.604.3023.2%
96033.913.199.0731.2%
96033.913.199.0731.2%

How much electricity does a -80 °C freezer use, and what does -70 °C save?

Across the 78 freezers on the list, the median daily use at -80 °C is 8.72 kWh, or 3,183 kWh a year. At -80 °C, daily use runs from 4.71 kWh to 13.19 kWh. Every model uses less at -70 °C. Larger freezers tend to use more. At -80 °C, freezers of 729 liters (25.74 cubic feet) and up, the larger half of the list, use a median of 9.43 kWh a day, and the smaller half uses 8.37 kWh.

What does a -80 °C freezer cost to run in each state?

At the 2025 U.S. average commercial price of 13.37 cents a kWh, the median daily use costs $426 a year at -80 °C and $292 at -70 °C. State prices run from 7.36 cents a kWh in North Dakota to 36.31 cents in Hawaii.

A year's electricity in the United States, at 13.37 cents a kWh (2025)
FreezerAt -80 °CAt -70 °C
Median use$426$292
Lowest use$230$167
Highest use$644$479

Each column takes the median, lowest and highest use at its own setting, so a row's two figures can come from different models.

To cost your own freezer, enter its daily use, your electricity price and your room's temperature in the operating cost calculator.

How much electricity do lab refrigerators and freezers use?

The list also holds 1,084 lab refrigerators and freezers that are not ultra-low temperature freezers. Most of them are in EPA's high performance class. The median high performance refrigerator uses 1.41 kWh a day, and the median high performance freezer 2.08 kWh. At close median volumes of 523 and 497 liters, high performance refrigerators with glass doors use more than those with solid doors, a median of 1.47 kWh a day against 1.15 kWh.

Daily energy use against volume, for the other refrigerators and freezers on EPA's list

Each dot is one model.

  • Doors
  • Glass
  • Solid
Each type's count and medians
Refrigerators and freezers on the list, by type and door
TypeDoorsModelsShare of the listMedian volume (liters)Median kWh a day
High performance refrigeratorsGlass72062%5231.47
High performance refrigeratorsSolid23220%4971.15
High performance freezersSolid11310%4762.08
General purpose refrigeratorsSolid101%7901.48
General purpose freezersGlass1under 1%1,3137.39
General purpose freezersSolid81%2291.42

Which refrigerants do lab freezers and refrigerators use?

Of the list's 1,162 models, 653 use propane (R-290), the most common refrigerant, and 506 use isobutane (R-600a). Ethane, R-170, is used only in ultra-low temperature freezers: 67 of the 78 use it. Every ultra-low temperature freezer on the list names two or more refrigerants, so the shares below add up to more than 100%. Only 1 model on the list uses R-404A, a blend with a global warming potential of 3,920. EPA gives propane 3 and ethane 5.5.

R-290Propane, global warming potential 3
653 models 56%
R-600aIsobutane, global warming potential 3
506 models 44%
R-170Ethane, global warming potential 5.5
67 models 6%
R-1150Ethylene, global warming potential 3.7
17 models 1%
R-50Methane, no value on either EPA page
2 models under 1%
R-704Helium, no value on either EPA page
2 models under 1%
R-404AA blend, global warming potential 3,920
1 model under 1%
R-601aIsopentane, no value on either EPA page
1 model under 1%

How much heat does a freezer put into the room?

A freezer at the median use gives off 363 watts at -80 °C, or 1,240 BTU an hour, and 249 watts at -70 °C, or 850 BTU an hour. That counts every kWh it uses as heat given off into the room, as the freezer room heat load calculator does.

To add up the heat of every unit in a room, open the freezer room heat load calculator.

Our ultra-low freezers and lab refrigerators

The energy figures above come from EPA's list, not from these units. Each card shows the unit's name and price from our catalog.

How we counted

Source
EPA's ENERGY STAR list of certified laboratory grade refrigerators and freezers, with its rows as updated on 13 November 2026 and read on 28 November 2026. This page leaves out the maker and model that the list names for each unit.
What the test measures
EPA's test runs each unit for a day in a room held at 24.0 °C, opens a freezer's door six times, and records the energy it uses. It tests each ultra-low temperature freezer at both -70 °C and -80 °C.
What it takes to be listed
To be listed, an ultra-low temperature freezer may use at most 0.46 kWh a day per cubic foot if it is under 20 cubic feet (566 liters). A larger one may use at most 0.35 kWh a day per cubic foot. The limits apply at -75 °C, between EPA's two test temperatures, and the list gives each freezer's own figure there.
Rows we left out or changed
We left out 1 of the list's 1,163 rows, a freezer listed with a cabinet temperature above 0 °C. Since only ultra-low temperature freezers get the -80 °C and -70 °C tests, we removed the -80 °C or -70 °C value the list gave 6 refrigerators.
Rows that repeat
The list gives each model number its own row, and some model numbers report the same test results. The 78 freezers report only 44 different results, so some pairs in the chart, and some rows in its table, match.
Volumes
EPA lists volume in cubic feet. We show liters, at 28.3168 liters to the cubic foot.
Electricity prices
The U.S. Energy Information Administration's average retail price of electricity for the commercial sector in 2025, by state, read on 28 November 2026. A year's cost is the daily use times 365 times the price.
Credit
Source: U.S. Energy Information Administration (Nov 2026)
Refrigerants
The global warming potentials come from two pages of EPA's Significant New Alternatives Policy program, one on very low temperature refrigeration and one on household refrigerators and freezers. R-50, R-601a and R-704 are on neither page, so they show none.
Your own freezer
A freezer in a lab can use more or less than its listing, because EPA's test fixes the room's temperature and the door openings, and a lab's room and habits differ.